Product Description:
The R&S ENV432 is a four-wire V-shaped artificial power network manufactured by R&S in Germany and distributed by EUTTEST. The ENV432 meets the requirements of CISPR 16-1-2, EN 55016-1-2 and ANSI C63.4 for V-shaped networks with an analog impedance of (50 μH + 5 Ω) in the frequency range from 9 kHz to 30 MHz. | 50 Ω with a Schuko socket (EUT 1) for the connection of the equipment under test (EUT) and a CEE 32 A socket (EUT 2).
R&S ENV432 productsFeatures:
- 9 kHz to 30 MHz frequency range
- Power Handling Capacity (up to 32 A), Constant Current
- Analog impedance (50 μH + 5 Ω) || 50 Ω according to CISPR 16-1-2
- V-networks according to CISPR, EN, VDE, ANSI, FCC Part 15 and MIL-STD-461D, E and F
- Calibrated according to CISPR 16-1-2 and ANSI C63.4
R&S ENV432Product Advantage:
- Empty core design as well as analog hand
- The R&S ENV432 4-wire V-network is based on an air-core inductor and contains an analog hand.
- Built-in 10 dB attenuator
- In order to ensure a standard impedance (without taking into account the input fading of the measuring receiver), the
- Equipped with a 10 dB attenuator.
- Built-in pulse limiter (can be turned off)
- A built-in pulse limiter (can be turned off) protects the measurement receiver input.
- Automatic temperature monitoring euttest
- The fan of the R&S ENV432 switches on automatically when the internal temperature of the housing reaches approximately +50 °C. This protects the V-network from high constant current loads. This protects the V-network from high constant current loads.
- Remote control of TTL levels (compatible with Rohde & Schwarz measurement receivers)
- TTL control inputs are available through the controller orRohde & Schwarz EMI Measurement ReceiverDrive for remote control of phase selection in automated test systems.
R&S ENV432 Technical Data::
<<<<提醒:左右滑动表格>>>>norm | ||
Frequency range | 9 kHz to 30 MHz | |
Analog Impedance | 50 μH + 5 Ω) || 50 Ω | |
Error limits according to CISPR 16-1-2 | Amplitude and phase | ±20% and ±11.5°. |
Decoupling fading between the power supply and the measurement receiver port according to CISPR 16-1-2 | 9 kHz to 50 kHz | > 0 dB to >40 dB (linear increase in logarithmic frequency) |
50 kHz to 30 MHz | > 40 dB | |
Test path to the EUT | ||
Maximum allowable constant current | EUT 1 | 16 A |
Supply Voltage | EUT 1 | 0 V to 240 V AC + 10% |
dc voltage | EUT 1 | 0 V to 350 V DC + 10% |
Maximum allowable constant current | EUT 2 | 32 A |
Peak current (transient) | EUT 2 | 50 A (2 minutes) |
Supply Voltage | EUT 2 | 0 V to 240/415 V AC + 10% |
dc voltage | EUT 2 | 0 V to 350 V DC + 10% |
power frequency | 0 Hz to 60 Hz + 5% | |
Test path to the measurement receiver | ||
Maximum Allowable EUT RF Nuisance Power | 5 W | |
Voltage division factor between EUT and measurement receiver port | Built-in attenuator, supplied calibration data (using V-network) | 10 dB - 0.5 dB/+ 2.0 dB |
Response threshold for built-in pulse limiter | closeable | 140 dB (μV) (nominal) |
Fan power and control logic | ||
Supply Voltage | 115 V setting | 100 V to 120 V AC ± 10% |
230 V setting | 220 V to 240 V AC ± 10% | |
power frequency | 50 Hz to 60 Hz ± 5% | |
power wastage | 100 VA (nominal) | |
connector | ||
Power supply and DC voltage output | Front Panel, EUT 1 | Schuko socket euttest |
Power supply and DC voltage output | Front Panel, EUT 2 | CEE socket (6 h) |
RF output | Front Panel, to Test Receiver | Negative, 50 Ω |
analog hand | front panel | 4 mm Fitting, Female Fitting, Clamp with Flange |
Power supply and DC voltage input | Rear Panel, Power Supply | CEE Built-in connector (6 h) |
Supply voltage input (auxiliary voltage) | Rear panel for fan and power supply for remote control | Low Temperature Connectors with Power Filters |
Remote Control Input | Rear Panel, Remote Control | 25 Contacts, D-Sub, Female Connectors |
Safety grounding wire | Front and rear panels | M8 Threaded Pins |
RF Reference Ground | both sides | Ground rod with seven M6 threads euttest |
Common data | ||
Operating Temperature Range | +5°C to +45°C | |
Storage temperature range | -40°C to +70°C | |
Overall size | W × H × D | 446 mm × 289 mm × 500 mm |
weight | 24 kg | |
electrical safety | Follow the instructions in the manual | According to EN 61010-1 |
EMC | According to IEC/EN 61326-1 | |
emit (a particle) | Category B, in line with the requirements of the residential environment | |
anti-interference | Meets the requirements of industrial environments |