Product Description:
The R&S ENV216 2-wire V-shaped artificial power network meets the requirements of CISPR 16-1-2 and EN 55016-1-2 for V-shaped networks with an analog impedance of (50 μH + 5 Ω ) || 50 Ω over the frequency range of 9 kHz to 30 MHz, and meets the relevant requirements of ANSI C63.4 for the frequency range of 150 kHz to 30 MHz. 150 kHz to 30 MHz. negative connector connection EUTs are available with ground protection contacts. several models of the ENV216 are available to comply with country-specific power connector systems. the ENV216 is designed for CE interference voltage measurements on single-phase EUTs.
Product Features:
- 9 kHz to 30 MHz frequency range
- Power handling capacity up to 16 A, constant current (varies by country)
- Analog Impedance (50 μH + 5 Ω) || 50 Ω per CISPR 16-1-2 Amd. 2:2006
- V-networks compliant with CISPR, EN, VDE, ANSI, FCC Part 15 and MIL-STD-461D, E and F
- Calibrated according to CISPR 16-1-2
Product Advantage:
- Empty core design as well as analog hand
- The R&S ENV216 2-wire V-network is based on an air-core inductor and contains an analog hand.
- Selectable switchable 150 kHz high pass filter
- To prevent the measurement receiver from being overloaded by high levels of the low-frequency spectrum, a 150 kHz high-pass filter can be inserted.
- Built-in 10 dB attenuator
- To ensure standard impedance (regardless of receiver input attenuation), the R&S ENV216 is equipped with a 10 dB attenuator.
- Built-in pulse limiter (can be turned off)
- A built-in pulse limiter (which can be turned off) protects the receiver input.
- Remote Control TTL Levels (with Rohde & Schwarz)EMI Measurement Receiver(Compatible)
- The TTL control inputs are driven via a controller as well as a Rohde & Schwarz measurement receiver and can be used for remote control (phase selection and activation of high-pass filters) in automated test systems. An optocoupler protects against external interference.
R&S ENV216 Technical data.
<<<<提醒:左右滑动表格>>>>Frequency range | 9 kHz to 30 MHz | |
Analog Impedance | Phase and amplitude | (50µH + 5Ω) || 50Ω |
error limit | ±20% (amplitude), ±11.5° (phase), compliant with CISPR 16-1-2 | |
isolation 1) | 9 kHz to 50 kHz | >0 dB to 40 dB (linear increase in logarithm of frequency) |
50 kHz to 30 MHz | >40 dB | |
Test path to the EUT euttest | ||
AC supply voltage | If the EUT is operated with AC voltages up to 90 V and DC voltages up to 50 V, the logic circuitry must be powered through the external plug-in power supply that comes standard. | 0 V to 240 V AC + 10% |
AC power frequency | 50 Hz to 60 Hz ± 5% | |
dc voltage | 0 V to 50 VDC | |
Maximum allowable continuous current | Varies by country/region | |
Model 12 | 16 A | |
Model 13 | 13 A | |
Model 14 | 16 A | |
Model 15 | 10 A | |
Model 16 | 15 A | |
Test path to the measurement receiver | ||
Maximum Allowable EUT RF Interference Power | 1 W | |
High-pass filter | Integrated, Selectable Switch | 150 kHz |
Voltage dividing factor between EUT and measurement receiver ports | Built-in attenuator with calibration data (using V-network) | 10 dB |
Response threshold for built-in pulse limiter | closeable | 140 dB (µV) |
terminal (electrical engineering) | ||
AC power input | Varies by country/region | Plug with grounding contact and 1.8 m cable |
AC supply voltage input for EUT | Varies by country/region | Receptacle with grounding contact |
RF output | Negative, 50Ω | |
Remote Control Input | 25 Contacts, D-Sub, Negative | |
Analog hand input | 4 mm Fitting, euttest, negative, with knurled fixture | |
reference ground | Ground rod with three M4 threads | |
grounding | M6 Threaded Pins | |
External power connector | Located on the rear panel and powered by the supplied plug-in power supply | DC hollow connectors Ø5.5 mm, 2.1 mm, 10 V to 18 VDC, 250 mA |
Common data | ||
Operating Temperature Range | +5°C to +45°C | |
Storage temperature range | -40°C to +70°C | |
size | W × H × D | 219 mm × 147 mm × 350 mm (8.6 in × 5.8 in × 13.8 in) |
weight | 5.5 kg (12.1 lb) | |
electrical safety | Follow the instructions in the manual | Conforms to EN 61010 |
EMC | Conforms to IEC/EN 61326 | |
radiate | Category B, in line with the requirements of the residential environment | |
anti-interference | Meets the requirements of industrial environments |