Introduction:
The ICI-DP HH500-15 is an IC double-pulse fault injection magnetic field probe manufactured by Langer-EMV, Germany, with a tip diameter of 500 μm. The double-pulse magnetic field source ICI-DP HH500-15 injects fast transient magnetic field pulses into DUTs (e.g. ICs), and is specifically designed for EM Fault Injection (EMFI) in IC safety applications. EMFI) in IC safety applications. Using this probe, single pulses as well as double pulse sequences with a minimum pulse following time of 25 ns can be injected into ICs with precise timing and position.
With the "Sync" input, single or double pulses can be injected into the DUT's function sequence in a synchronized manner. The probe is powered and controlled by the pulse generator BPS 204.
Technical parameters:
<<<<提醒:左右滑动表格>>>>Probe Model | ICI-DP HH500-15 |
use | IC fault injection of magnetic field pulses |
manufacturer | Langer emv |
agent | Euttest |
Probe tip diameter | 500μm |
Maximum pulse injection voltage | 1000V |
rising time | 2ns |
Maximum current | 16A |
pulse width | 10ns |
Minimum Pulse Sequence Time | 25ns |
repetition rate | 0.1 Hz - 15 kHz (single pulse, <500 V) 0.1 Hz - 7.5 kHz (double pulse, <500 V) |
polarities | bipolar |
Minimum trigger pulse delay | 35ns |
Accessory Trigger Pulse Delay | Controlled by BPS 204 pulse generator |
Probe power supply | Powered by BPS 204 |
quality | 190g |
size | 26 mm x 54 mm x 71 mm |