What is HI-6153?
ETS-Lindgren's Model HI-6153 Laser-Driven Electric Field Probe provides broadband EMF frequency coverage and a wide dynamic range to meet most test requirements. To take advantage of this capability, the HI-6153 RF Field Probe is designed for single-range readout so that data can be read continuously over the entire dynamic range. Data values for each axis (X, Y, Z) can be read individually or summed.
HI-6153 Test Method:
Testers can connect the Model HI-6153 probe to the Model HI-6100 Field Monitor using fiber optic signal and power cables, or directly to PC USB Port™ laser software with the HI-6113 laser data interface and ProbeView.
The HI-6100 provides manual functions and programmed control via IEEE-488 and RS-232 serial data interfaces. Up to four probe readings can be displayed simultaneously.
E-Commerce Center™ Modular RF Platform and EMSense™ In addition to functioning as a system-level platform, the interface card can be used as a field monitor with the HI-6000 series field probes.
The HI-6113 provides laser power and communications for the HI-6153. A USB connection to a PC allows for quick and easy data collection using the ProbeView laser software.
How to buy HI-6153?
EUTTEST Shenzhen EUTTEST Technology Co., Ltd. is the agent of ETS-LINDGREN, and can provide consulting service for all the products price and delivery period of ETS company.
2023: HI-6153 Discontinued. Please select the battery-powered HI-6053 below:
HI-6053 Battery Powered Electric Field Probe Frequency Range 10 MHz-40 GHz
The main technical parameters of HI-6153:
<<<<提醒:左右滑动表格>>>>Model number | HI-6153 |
manufacturer | ETS Lindgren |
Agents in China | Shenzhen EUTTEST |
Sensor type | electric field |
sensing | Isotropic (X, Y, Z axis readings) |
Frequency range | 10MHz-40GHz |
accurate | ±2.5dB (10MHz-18GHz), +2.0dB/-4.0dB (18-40GHz) |
dynamic range | 2.0-800V/m |
resolution (of a photo) | 0.01V/m |
concentric deviation | ±1.0dB |
Readout units | V/m, V2/m2mW/cm2 |
biometrics | ±0.5dB |
sampling rate | >70 times per second |
overloaded | >1500V/m CW |
optical interface | Dual Fiber (62.5μm Multimode), FC Interface |
operating temperature | 10 - 40°C |
relative humidity | 5%-95% non-condensing |
size | 43.2cmx10.2cm |
weight | 540g |
Charts related to technical parameters:

Frequency Response Curve of HI-6153