Introduction to F-33-1:
The F-33-1 is manufactured by FCC in the U.S.A. and licensed to EUTTEST for sale in China, it is used for laboratory and field testing, the probe has a small outside diameter of about 71 mm and an inside diameter of 32 mm. the typical transmission impedance of the F-33-1 is 15 dB (±2 dB) over the frequency range of 70 kHz to 250 MHz. RF currents range from 10 amps continuous wave and 100 amps peak with a duty cycle of 0.1. Maximum current on the primary power line is 100 amps (from DC to 400 Hz).
The main technical parameters of the F-33-1:
Available frequency range: 10 kHz-250MHz;
Internal ring diameter: 32 mm;
External ring diameter: 71 mm;;
Height: 19mm;
Typical value ZtΩ(1): 6.3;
Typical dBΩ: 16 ;
Connector: Type-N, N female, 50 ohms;
Rated power current value: 100A (DC- 400 Hz);
RF (CW continuous wave): 2A;
Peak pulse current (2): 50A;
1: Calibrate the probe with a 50Ω+j0Ω load impedance.
2: Depends on pulse width and pulse repetition rate.
Matching Calibration Test Fixtures for F-33-1.
fcc-mpcf-3-32/71/19
Packing list for F-33-1:
One F-33-1; one instruction manual;
One copy of factory calibration data; one hard plastic case;
FCC: Product specifications are subject to change without notice.