Introduction:
CDN M3PE 16A 1000V is a 3-wire coupling and decoupling network manufactured by SCHWARZBECK Company in Germany, authorized by Shenzhen EYT Company for sales and quotation service. This CDN coupling and decoupling network supports DUT line-to-ground voltage up to 500V AC and 1000V DC, and maximum saturation current up to 16A, which can be used for EMC testers to conduct conducted common-mode immunity test according to IEC 61000-4-6 Annex D.2, and is suitable for conducting immunity test along the power line of unshielded AC and DC power lines (utility). It is suitable for conducting immunity tests along power lines for unshielded AC and DC power lines (mains).
Both this model and the CDN M2M3PE 16A have 16A pass-through current, but the CDN M3PE described on this page is slightly less powerful in that it only supports the CDN-M3 test mode and cannot be switched to CDN-M2 test use. If you want to test M2 or M3 mode, please purchase the CDN M2M3PE 16A The
Technical parameters:
<<<<提醒:左右滑动表格>>>>Model number | CDN M3PE 16A 1000V |
Device name | Three-wire coupled decoupling network-CDN |
manufacturer | Schwarzbeck |
agent | EUT TEST |
Test frequency range | 150kHz-230MHz |
Common mode impedance of CDN | At 150 kHz - 24 MHz: 150 Ω ±20 Ω At 24MHz - 80MHz: 150 Ω +60 Ω / -45 Ω 80 MHz - 230 MHz: 150 Ω ±60 Ω See common mode impedance graph below for detailed values. |
Maximum RF Voltage | 30V |
Maximum input power | 6W |
connector | BNC |
impedance | 50Ω |
Maximum Operating Voltage - Line to Ground | 500V AC 1000V DC |
Maximum operating voltage - line to line | 1000V AC 2000V DC |
saturation current | 16A |
size | 216 mm*105 mm*108 mm |
weight | 1.4 kg |
Recommended Calibration Options | CA 3/4 SR 30/4 SR 100-6W |
Common mode impedance plot:
The following figure shows the common mode impedance curve of this product in accordance with the requirements of IEC 61000-4-6 and GBT 17626.6 conducted immunity test standard, the red line in the figure represents the impedance value range given by the standard.

Common mode impedance graph
Product Recommendation:
- Learn how to shop for a CDN?
- Need CDNE?
- Need more CDN & CDNE Coupling & Decoupling Networks models with test voltages and number of cables under test?
Please go to the following link to view: